7-3
IM 253710-01E
Setting the Trigger
7
7.2
Selecting the Trigger Source
«For a functional description, see section 1.4.»
This section applies when the trigger mode is set to a mode other than OFF.
Keys
• The
mark indicates the keys that are used for the operation.
• To exit the menu during operation, press the ESC key.
CONFIGURATION
ELEMENTS
VOLTAGE
CURRENT
FILTER
1
FILTER
FILTER
FILTER
2
3
4
RES
ET SELECT
CH
8
CH
6
CH
4
CH
2
CH
7
CH
5
CH
3
CH
1
OBSERVATION TIME
TRIG’ D
ACQ
REMOTE
MENU
SHIFT
NULL
CAL
COPY
LOCAL
HELP
MISC
FILE
ZOOM
CURSOR
MATH
DISPLAY
SETUP INPUT MEASURE TRIGGER
ABORT
E S C
SINGLE
START
START/STOP
Procedure
Set the trigger mode to a mode other than OFF. For procedures, see section
7.1, “Selecting the Trigger Mode.”
1.
Press the TRIGGER key to display the Trigger setting menu. Check that [Mode]
is set to a mode other than [OFF].
2.
Press the [Source] soft key to display the source selection box.
3.
Turn the jog shuttle to select the source from [CH1] to [Ext].
4.
Press the SELECT key to confirm the new source.