15-16
IM 253710-01E
15.6 Setting the Motor’s Number of Poles Used to
Compute the Synchronous Speed and the Slip
Keys
• The
mark indicates the keys that are used for the operation.
• To exit the menu during operation, press the ESC key.
CONFIGURATION
ELEMENTS
VOLTAGE
CURRENT
FILTER
1
FILTER
FILTER
FILTER
2
3
4
RES
ET SELECT
CH
8
CH
6
CH
4
CH
2
CH
7
CH
5
CH
3
CH
1
OBSERVATION TIME
TRIG’ D
ACQ
REMOTE
MENU
SHIFT
NULL
CAL
COPY
LOCAL
HELP
MISC
FILE
ZOOM
CURSOR
MATH
DISPLAY
SETUP INPUT MEASURE TRIGGER
ABORT
E S C
SINGLE
START
START/STOP
Procedure
1.
Press the INPUT key to display the Input setting menu.
2.
Press the [Motor Module] soft key to display the motor module setting dialog
box.
On products (PZ4000) with firmware version before 2.01, only the [Power Module] soft
key appears and you cannot set the motor module. In addition, if the motor module is
not installed in the element number 4 slot, the [Motor Module] menu does not appear.
Setting the motor's number of poles
3.
Turn the jog shuttle to select [Pole].
4.
Press the SELECT key to display the motor’s number of poles selection box.
5.
Turn the jog shuttle to set the number of poles.
For the procedures related to entering values using the jog shuttle, see section 4.1,
“Entering Values and Strings.”
6.
Press the SELECT key or the ESC key to close the setting box.