15-8
IM 253710-01E
Selecting the input range of the torque meter signal
Two types of range settings are available: fixed range and auto range.
• Fixed range
Select from the following input ranges:
50Vpk, 20Vpk, 10Vpk, 5Vpk, 2Vpk, and 1Vpk.
• Auto range
Select [Auto] for the input range setting to enable auto range. The range changes
automatically depending on the amplitude of the input signal. The switching
conditions and precautions are the same as the descriptions given in section 5.3,
“Setting the Measurement Range during Direct Input.” The different ranges used in
the auto mode are the same as those available for fixed range.
15.2 Setting the Input Range of Revolution Sensor and Torque Meter Signals