2-7
SETUP
Instrument Settings
There are five major instrument settings that you can change. They are:
•
System Setup
•
Date and Time
•
Hardware Setup
•
NetProfiler Setup and Status
•
Language
Choosing System Setup Options
1. Select
OPTIONS
from the ColorEye XTH
Main Menu
.
2. Select
SETTINGS
from the
Options
Menu
.
3. Select
SYSTEM SETUP
from the
Settings Menu
.
4. The System Setup Screen is shown below:
5. To change an item in the System Setup list:
a. Move the cursor to the desired item using the left or right arrow keys.
b. Press
Enter
to scroll through the available values for each item.
6. Repeat step 5 for all the items you need to choose in the System Setup.
7. Save the new System Setup by moving the cursor to the
[save]
at the bottom of
the screen. When the
[save]
(lowercase) changes to
[SAVE]
(uppercase), press
Enter
.
Description of Options on the System Setup Menu
The System Setup Menu contains your preferences for how the instrument operates
when taking measurements. For example, What photometric range should be used?
Should measurements be automatically saved (Auto Saving option)? If so, what should
they be called (Auto Naming option)? The System Setup items and their function are
described below:
Color Ratio
Variable:
If CMC is chosen, then 1:1 is appropriate for determining perceptible color difference,
while 2:1 is appropriate for determining acceptable color difference. If CIE94 is chosen,
then 1:1:1 is appropriate for determining perceptible color difference, while 2:1:1 is
appropriate for determining acceptable color difference.
Photo(metric) Range
:
120%
or
200%
.
If your sample is optically brightened, choose 200%. If it is not optically brightened or is
only slightly optically brightened, choose 120%.
Std Avg.
:
1
through
9
.
You can set up your instrument to average measurements to achieve a standard. You
must indicate the number of standard measurements to average.
SYSTEM SETUP:
COLOR RATIO
1:1
PHOTO RANGE:
120%
STD AVG.
1
TRIAL AVG.
1
CALIB AVG.
3
AUTO PRINT
OFF
MI 2nd ILLUM
C
AUTO STANDARD
OFF
AUTO NAMING
OFF
AUTO SAVING
OFF
ASSOCIATIONS
NO
TOLERANCES
SYS
[save]
>
Содержание COLOREYE XTH
Страница 1: ...COLOREYE XTH Spectrophotometer OPERATION MANUAL ...
Страница 20: ...INSTALLATION 1 16 ...
Страница 46: ...OPERATION 3 6 ...
Страница 54: ...MAINTENANCE 4 8 ...
Страница 57: ...A 3 APPENDIX Color Check Record Temperature Aperture Size Serial Number of Instrument Date L a b ...
Страница 58: ...APPENDIX A 4 ...
Страница 62: ......
Страница 63: ......