1-15
INSTALLATION
General
Illumination . . . . . . . . . . . . . . . . . . . . . . Pulsed Xenon D65
Spectral Range . . . . . . . . . . . . . . . . . . . 360 nm to 750 nm
Wavelength Interval. . . . . . . . . . . . . . . . 10 nm
Photometric Range (selectable) . . . . . . 0%-120%, reflectance
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0% - 200%, reflectance
Photometric Resolution . . . . . . . . . . . . . 0.01%
Baud Rate . . . . . . . . . . . . . . . . . . . . . . . 2400 - 38400
Measurement Cycle Time . . . . . . . . . . . 1 sec.
Optical Geometry Configuration . . . . . . D/8 (diffuse)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Simultaneous Specular Component
Excl./Specular Component Incl.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38mm diameter integrating sphere,
Dual Beam
Aperture Size
Regular Area of View (standard)
Illuminated/Measured . . . . . . . . . . . . . . 10mm circular diameter/
5mm circular diameter
Small Area of View (optional) . . . . . . . .
Illuminated/Measured . . . . . . . . . . . . . . 5mm circular diameter/
2mm circular diameter
Repeatability
Typical: 0.05 RMS Delta E CIELAB
Maximum: 0.10 RMS Delta E, CIELAB (Refer to Note 1 below.)
Interinstrument Agreement
Typical: 0.15 Avg. Delta E CIELAB (Refer to Note 2 below.)
Maximum: 0.25 Average Delta E, CIELAB (Refer to Note 2 below.)
Notes:
1.
Representative of all instruments’ performance using a stable
white tile under controlled laboratory conditions.
2.
Representative of all instruments’ performance using the
average from the norms of 13 BCRA color tiles under
controlled laboratory conditions.
Representative values (described in NOTES 1 and 2) may vary with the type(s) of
samples measured.
Service
a. Service Properly—With the exception of those procedures given in the
Maintenance chapter of this manual, do not attempt to service this
product yourself. If you should attempt unauthorized repairs yourself,
you may invalidate the warranty.
b. Call
X-Rite
when You Have Questions— A full complement of
service and applications experts are ready to deliver the finest customer
support and service available in the industry. When you contact
X-Rite,
you contact a group of professionals who are totally
dedicated to your individual satisfaction in products and services. Thank
you for buying a
X-Rite
product. We appreciate your business
and look forward to working with you.
Содержание COLOREYE XTH
Страница 1: ...COLOREYE XTH Spectrophotometer OPERATION MANUAL ...
Страница 20: ...INSTALLATION 1 16 ...
Страница 46: ...OPERATION 3 6 ...
Страница 54: ...MAINTENANCE 4 8 ...
Страница 57: ...A 3 APPENDIX Color Check Record Temperature Aperture Size Serial Number of Instrument Date L a b ...
Страница 58: ...APPENDIX A 4 ...
Страница 62: ......
Страница 63: ......