Ionselective measurement
Ionselective measurement
92
BA61101/08.00/IWa/MLP5-E2
BA61101/08.00/IWa/MLP5-E2
93
Ã
ISE CAL
25.4°C
Ê
C1= 11. 0
Enter the concentration of
the first calibration standard
solution.
: Select figure
: Set figure.
Rinse electrode and temperature probe and immerse
into first standard solution.
ÃC2
ISE CAL
U1= 10.8mV
25.4°C
C1= 11.20
Start measurement. Wait
for stable test values.
Ã
ISE CAL
25.4°C
C2= 00.00
É
Enter the concentration of
the second calibration
standard solution.
: Select figure
: Set figure.
Rinse electrode and immerse into the second
standard solution.
ÃCalDat
ISE CAL
U2= -45.3mV
25.4°C
C2= 102.10
Start measurement. Wait
for stable test values.
U0= -102.3mV
ÂISE
S= -57.5mV/D
ISE CAL
Display of calibration data.
ÂISE
ISE CAL
C1= 11.20
C2= 102.10
Indication of used standard
solutions.
À
and
Á
switch to-and-fro between
the calibration data.
Ãmeth.
Âmenu
ÁISE
25.4°C
TP
ÀCal
Back to ISE menu.
Any ionselective measurement
Ãmeth.
Âmenu
ÀCal
ÁpH
25.4°C
AR TP
Position 10: pH menu.
mV
Âmeas.
14697 Äa-Ten
¡pH
Method change.
14779 ÅHS
Âmeas.
ISE F
¡ISE
ISE
= Select ionselective
measurement.
Ãmeth.
Âmenu
ÁISE
25.4°C
TP
ÀCal
Entry to ISE.
Âmenu
ISE
25.4°C
TP
35,7
Start measurement.
ISE measurement runs
continuously.