Troubleshooting
Troubleshooting
218
BA61101/08.00/Pla/MLP5-E2
BA61101/08.00/Pla/MLP5-E2
219
pH
time-out
Âmenu
ÀCal
ÁpH
25.4°C
AR TP
Measurement: Measured value not
stable.
Possible cause / Action:
- Junction of the electrode contaminated
/ Clean junction
- Unsuitable electrode
/ Change electrode
- Sample solution not stable
/ Depends on aim of the measurement
and the characteristics of the sample
solution
- Electrode too old / Change electrode
- Temperatures of sample solutions too
different
/ Wait for temperature adjustment
time-out
CalErr AR TP
ÀCal
Calibration: Measured value not stable.
Possible cause / Action:
- Junction of the electrode contaminated
/ Clean junction
- Cable broken
/ Change cable or electrode
- Electrode too old
/ Change electrode
CalErr AR TP
ÀCal
Calibration: Calibration data outside of
admissible ranges.
Admissible slope range:
-62 mV ... -50 mV
Possible cause / Action:
- Wrong buffer solutions
/ Change buffer solutions, see
set buffer combination
- Old buffer solutions / Change buffer
solution, see preservability date
- Old electrode / Change electrode
- Short circuit / Change electrode or
repair meter
- Wrong temperature value (only with
measurements without temperature
probe)
/ Set correct temperature value
Admissible asymmetry range:
-30 mV ... +30 mV
Possible cause / Action:
- Wrong buffer solutions
/ Change buffer solutions, see
set buffer combination
- Old buffer solutions / Change buffer
solution, see preservability date
- Old electrode / Change electrode
- Contaminated junction
/ Clean junction
- Wrong temperature value (only with
measurements without temperature
probe)
/ Set correct temperature value