Conductivity undependent on the position of selector switch!
Command Function
CLF1A
Switch on AutoRange, call up conductivity- and
temperature meas. value
Cond 22.02.94 09:00:28 0.0 uS/cm 21.9 °C
MLP5 >
CLF12
Meas. range 0.000...3.000 µS (only when C<0.15/cm)
Cond 22.02.94 09:00:28 0.000 uS/cm 21.9 °C
MLP5 >
INVALID COMMAND
when C > 0.15/cm
CLF13
Meas. range 0.00...30.00 µS
Cond 22.02.94 09:00:28 0.00 uS/cm 21.9 °C
MLP5 >
CLF14
Meas. range 0.0...300.0 µS
Cond 22.02.94 09:00:28 0.0 uS/cm 21.9 °C
MLP5 >
CLF15
Meas. range 0...3000 µS
Cond 22.02.94 09:00:28 0 uS/cm 21.9 °C
MLP5 >
CLF16
Meas. range 0.00...30.00 mS
Cond 22.02.94 09:00:28 0.00 mS/cm 21.9 °C
MLP5 >
CLF17
Meas. range 0.0...300.0 mS
Cond 22.02.94 09:00:28 0.0 mS/cm 21.9 °C
MLP5 >
CSAL
Meas. range 0.0...70.0
Sal 22.02.94 09:00:28 0.0 21.9 °C
MLP5 >
RC
Read-out of cell constant
0.690/cm
MLP5 >
RTC
Read-out of temperature coefficient
nlF
MLP5 >
RREF
Read-out of reference temperature
25°C
MLP5 >