Conductivity measurement
Documentation
98
BA61101/08.00/IWa/MLP5-E2
BA61101/08.00/IWa/MLP5-E2
155
Preparation
Measurement settings required:
1.
Cell constant C
(dependent on the type of the meas. cell;
required after change of meas. cell).
2.
Temperature function TC (dependent on type of
measuring solution)
3.
Reference temperature: 25°C (recommended acc. to
DIN 38404 part 4)
Measurement
solution type
Temperature function
TC
Reference
temperature
natural water
(ground-, surface-
and drinking water)
set
nlF
to be set
distilled water
set
nlF
to be set
Other aqueous
solutions
determine
to be set
Salinity (seawater)
automatic nlF
acc. to IOT
*
automatic 20°C
non-aqueous
solutions
determine
to be set
*
International Oceanographic Table
(UNESCO Technical Papers in marine sciences 36)
Storage
To store results in memory for later documentation.
Example: Storage of data at site in the MultiLab P5 and output to a
printer or PC in the office.
- Max. 1000 results can be stored, in case of a higher quantity
of measuring values the last 1000 values remain in the memory.
- Free storage locations are displayed in a descending
sequence (1000, 999, 998.. 2, 1), after 1000 stored values the
number starts again from the beginning (..2, 1, 1000, 999, 998..).
- The Numbering of documented test values (Number, Sample
location) e.g. for the printout runs in ascending sequence
1 to 999, for control the functions "no. cont." or "number new".
¡
Function indication
form selected.
(Selector switch set e.g. to
position 2,
test C2/25 COD).
¡
Select function
Documentation.
¡
Preset:
"no" documentation.