Select funtion
documentation "storage".
¡
Function "Numbers"
selected.
Storage of test values.
¡
Function "no. cont 001"
set.
Test values continuously
numbered from 001 to 999.
001: No current test value
e.g. after RESET or erasure
of test value numbers (s.b.).
112: Next documented test
value
(e.g. no. 112)
after already performed
measurements with set
documentation.
In the test value memory always the last 1000 test-
values are available, not dependent on the current
number of the documented test values.
¡
Set function "number
new".
Connectable measuring cells
Recommended meas. cells:
Description
Model
o.-no.
Standard-conductivity meas.cell.
4-electrode-meas. cell with
integrated temperature sensor
TetraCon
96
301 244
Instrument meas. range
Cell constant cm
-1
3 µS/cm... 300 mS/cm
0.609
±
1.5%
Description
Model
o.-no.
Conductivity meas. cell with
integrated temperature sensor
and flow through beaker
LR 01/T
302 520
Instrument meas. range
Cell constant cm
-1
0.001 µS/cm... 300 µS/cm
0.1
±
2%
Description
Model
o.-no.
Trace-conductivity meas. cell
with integrated temp.sensor
LR001/T
302 530
Instrument meas. range
Cell constant cm
-1
0.001 µS/cm... 30 µS/cm
0.01
±
2%
Description
Model
o.-no.
4-electrode conductivity meas.
cell, glass/platinum with
integrated temp. sensor
TetraCon
L
301 255
Instrument meas. range
Cell constant cm
-1
50 µS/cm... 300 mS/cm
approx. 1