Configuration
Configuration
198
BA61101/08.00/IWA/MLP5-E2
BA61101/08.00/IWA/MLP5-E2
199
Âmenu
ÁTest1 m¥/L
Ãmeth. dH
Back to function
"Measurement".
Printout of own method:
(Example)
MultiLab P5
Software: A 28.03.94
04.01.15 16:07:00
Test1
445nm
2.0 - 20.0 [mg/L]
NH4-N
E0= 0.022000
S= 0.046000
14mm
Filter 12
Input after installation of a 12
th
filter. Input regardless of actual filter
position.
Âmeas.
¡language
date/time
system info
Menu "Configuration".
Âmeas.
¡user meth
system info
sequence
¡
Select function
"User meth".
Âconfigure
¡data
lin.re¥.
user meth
¡
Characteristic data
selected.
Âconfigure
¡filter 12
data
user meth
¡
Select Filter 12.
Âconfigure
¡new
erase
filter 12
¡
Filter 12 new selected.
Âconfigure
3 0nm
Ì
filter 12
¡
Input filter 12
wavelength.