Photometry - Additional functions
Photometry - Additional functions
52
BA61101/08.00/Pla/MLP5-E2
BA61101/08.00/Pla/MLP5-E2
53
Insert vial/cuvette with
sample:
Test value is displayed.
Absorbance
measurement:
A
beside the test value.
Back to measurement.
Measurement of absorbance against a reference value
The measurement of absorbance against a reference value considers
disturbing influences of colorations or turbidities of the sample.
The measurement of the reference value is performed before
measurement of the sample. To erase the reference value quit the
procedure or switch off the instrument.
¡
Function Absorbance
selected.
Adjusted wavelength
e.g. 690 nm.
Insert vial/cuvette with
reference solution :
Absorbance value is
displayed.
Measurement of
absorbance as reference
value Eref.
The reference value is
e.g. 0.132 E and is
stored.
Insert vial/cuvette with test
solution.
Measurement of
absorbance of the sample.
Display of corrected
absorbance of the sample.
¢
= Reference value is
subtracted from test value.
Insert further vials/cuvettes
with test solutions.
Measurement of
absorbance of the sample.
Display of corrected
absorbance of the sample.
The correction is carried out
with the last stored
reference value.
¢
= Reference value is
subtracted from test value.