Technical data
Technical data
240
BA61101/08.00/IPro/MLP5-E3
BA61101/08.00/IWa/MLP5-E2
241
Temperature
Meas. range
Resolution
Accuracy temperature
(at 15°C..35°C ambient
temperature)
Sensor exchange error
Temperature range
TetraCon 96
-5.0°C..90°C
0.1K
±
0.1K
±
1 digit
<0.2 K at -5.0°C .. 60.0°C
<0.4 K at 60.0°C .. 110.0°C
0°C..90°C
Check
with 0.01 m KCl and adjustment of the
cell constant
Dissolved oxygen O
2
(T
1
, T
2
, p)
Parameters
mg/L O
2
, mbar pO
2
, % saturation,
temperature continuously
Meas. principle
potentiostatic 3-electrode system
Meas. ranges (resolution)
0.0 to 300.0 mg/L (0.1 mg/l)
0%..999 % (1%) saturation.
In the range 0.0 mg/L..30.0 mg/L the
resolution is selectable between 0.1
mg/L and 0.01 mg/L.
In the range 0.0%..100.0% the
resolution is selectable between 0.1%
and 1%.
Accuracy
0.5 % of meas. value
(0.01 mg/L resolution)
Temperature
compensation
only slow NTC with OxiCal
automatically with double thermistor
system
Air pressure correction
AutoBar
automatically from 850 to 1080 mbar
Salinity correction
0..99, manually adjustable
Temperature
Meas. range
Resolution
Accuracy temperature
(at 15°C..35°C ambient
temperature)
Sensor exchange error
Temperature range
TriOxmatic
300
-5.0°C..60°C
0.1K
±
0.1K
±
1 digit
<0.2K at -5.0°C .. 60.0
°
C
<0.4K at 60.0°C .. 110.0
°
C
0°C..50°C
Functions
AutoReg, membrane control, AutoBar,
AutoRead, OxiCal, H
2
O-Cal,
BSB5/BOD
Calibration DIN EN 25814
OxiCal with OxiCal P or OxiCal SL
H
2
O-Cal with air-saturated water.