Configuration
Configuration
192
BA61101/08.00/IWA/MLP5-E2
BA61101/08.00/IWA/MLP5-E2
193
Âconfigure
¡unit ©------
Ãcit.
user meth
Input max. 7-digit unit.
Âconfigure
¡unit m¥/ ---
Ãcit.
user meth
L
e. g. mg/L.
Âconfigure
¡cit. ©----
ÃÇC
user meth
Input 5 digit indication
form.
Âconfigure
¡cit. d
ÃÇC
user meth
H
e. g. dH.
Âconfigure
¡ÇC
Ãzero
user meth
È.É
Select resolution.
ÂKonfigurieren
¡ÇC
ÃNull
Meth.Benutzer
É
e. g. 1.
Âconfigure
¡C1= 0 00
ÃE1
user meth
È
¡
Input first
concentration.
Âconfigure
¡C1= 0100
ÃE1
user meth
È
¡
Input first concentration,
e.g. 100.
Âconfigure
¡A1= 0. 00 A
ÃC2
user meth
È
¡
Input first absorbance.
Âconfigure
¡A1= 0.100 A
ÃC2
user meth
È
¡
Input first absorbance,
e.g. 0.100.
Input further 9 pairs of values acc. to above schema.
Âconfigure
¡A10= 0. 00 A
ÃMBA
user meth
9
¡
Input last absorbance,
e.g. 0.900.
Âconfigure
¡MBA 000
ÃMBE
user meth
È
Input beginning of meas.
range. Quantity of the
indicated numbers depends
on the resolution.
Âconfigure
¡MBA 001
ÃMBE
user meth
È
e. g. 10.