Troubleshooting
Troubleshooting
214
BA61101/08.00/Pla/MLP5-E2
BA61101/08.00/Pla/MLP5-E2
215
Photometry
Measurement
Generally: Overflow
Cause / Action:
- Lamp defective / Change lamp
- AD converter overloaded / Dilute
sample
- AD converter defective / Send meter to
WTW for repair
Generally:
Zero adjustment missing.
Action:
Perform zero adjustment
(see page 208)!
Measurement:
Measured value is below the measuring
range.
The displayed measured value (e.g. 0.16
mg/L) is an assessed value.
Action:
- Give result as assessed value
- Use reagent set with lower measuring
range
- Select vial/cuvette with thicker layer.
Measurement:
Measured value exceeds measuring
range of the reagent set.
The displayed measured value (e.g.
11.07
mg/L) is an assessed value.
Action:
- Give result as assessed value
- Use reagent set with higher measuring
range
- Select vial/cuvette with thinner layer.
Measurement:
Measured value exceeds measuring
range of the reagent set.
It is not possible to compute an assessed
value.
Action:
- Pre-dilute sample
- Use reagent set with higher measuring
range
Measurement with blank value adjustment
Sample blank value adjustment:
Small influence of the sample blank value.
The sample blank value adjustment ist
only necessary for very precise
measurements or for measurements with
very low concentrations.