The data transmission of the own method is done in a string. The
different data blocks of the string are separated by spaces:
5 characters method indication
1 space
5 characters wave length
1 space
9 characters unit
1 space
12 characters indication form
1 space
12 characters zero point:
Exponential graph is possible.
admissible signs are: " + ", " - ", " . ", " e ",
and all numbers from 0 ... 9.
1 space
12 characters slope: refer to zero point.
1 space
12 characters beginning of meas. range
1 space
12 characters beginning of meas. range
1 space
2 characters reference cuvette
1 space
5 characters resolution
Example:
CEMETest1 690nm mmol/L NH4-N 0.987e-3 2.12345 0.1 22.3 14 0.1
The recognition of an error leads to error message "INVALID
COMMAND".
The following inputs lead to error message "INVALID COMMAND" as
well:
- Slope = 0 or > 32000
- Zero pointt > 32000
- Beginning of meas. range < 0 or > 32000
- End of meas. range < 0 or > 32000
- End of meas. range
£
Beginning of meas. range
- Ref. cuvette different 10 mm, 14 mm, 20 mm or 50 mm
- Wave length does not correspond to the instrument
- Resolution not 0.1, 0.01, ...
After successful input the instrument displays the received method :
Ãmeth. NH¿-N
Âmenu
ÁTest1
mmol/L
The instrument limits independently the meas.
range (no remote indication!), when the entered
method data allow extinction values > 3 resp. > 2,5
(at 340 nm).
Totally 50 own methods are possible, max. 40 per filter position.
The 51st own method on one filter causes "Meth. Error"; in this
case another method should be erased before input of the new
method.