pH measurement
pH measurement
68
BA61101/08.00/IWa/MLP5-E2
BA61101/08.00/IWa/MLP5-E2
69
Ã
: Two point calibration: Continue calibration (see below)
Rinse pH electrode and temperature probe and
immerse into second buffer solution.
à TP
ConCal SLO pH
T = 23.6°C
pH=
Ì.ÊÎ
Start measurement.
Wait for stable test values.
à TP
ConCal SLO pH
T = 23.6°C
pH=
Ì.ÈÉ
Adjust pH value for the
measured temperature.
ConCal
S= -59.2mV/pH
Uasy= -1 mV
ÂpH
Display of electrode data.
The data are stored with
date and time.
ÂpH
ConCal
ASY: 6.97
SLO: 4.01
Indication of pH values of
the used buffer solutions.
À
and
Á
switch to-and-fro
between the calibration
data.
Ãmeth.
Âmenu
ÀCal
ÁpH
25.4°C
AR TP
Back to pH menu.
Additional functions
Switching on/off the drift control AR (AutoRead)
The drift control AR can be switched on or off for the pH
measurement.
If the drift control is switched on
AR
is displayed.
Note:
The drift control AR is separately switchable for pH and oxygen-
measurement!
AutoRead AR:
Measuring period 30 s up to 10 minutes, max. drift < 0.1 mV/30s
Ãmeth.
Âmenu
ÀCal
ÁpH
25.4°C
AR TP
Position 10: pH menu.
dpH 0.1
ÂpH
documentation
yes
¡AR pH
Menu "additional functions"
of pH measurement:
yes: Drift control AR
selected.
dpH 0.1
ÂpH
documentation
¡AR pH
no
Switch off drift control.
no: Drift control AR
switched off.
Ãmeth.
Âmenu
ÀCal
ÁpH
25.4°C
TP
Back to pH menu.