Oxygen measurement
Oxygen measurement
130
BA61101/08.00/IWa/MLP5-E2
BA61101/08.00/IWa/MLP5-E2
131
Final measurement
When the final measurement is called up the numbers of all samples
ready for final measurement will appear.
Dependent on the selected bottle number the instrument
automatically changes to the corresponding sample characterization.
Possibly two measurements of dilution water have to be performed.
Endmessun¥
ÂBSB5
Anzei¥e
¡Anf.messun¥
BSB 5 menu.
Auswertun¥
ÂBSB5
Anf.messun¥
¡Endmessun¥
Select BSB 5 final
measurement
("Endmessung").
01.03.94/02
BSB5 Endmess.
¡ 01.03.94/01
02.03.94/01
Entry to BSB 5 final
measurement. Display of
samples for final
measurement.
01.03.94/03
BSB5 Endmess.
¡01.03.94/02
01.03.94/01
Select sample.
If there is no sample for final measurement, the following is
displayed:
Ã
BSB5 Endmess.
keine Werte
ÃMessen
BSB5 Endmess.
01.03.94/02
Fl.Nr. E 24
É
Instrument proposes a
bottle, e.g. E124.
Given number may be
changed.
Immerse probe into bottle no. E124.
AR
BSB5
E124
Oº
5,7
m¥/L 20.1°C
Start measurement for e.g.
bottle E124.
à ¡AR
BSB5
ÁE124
Oº
5,6
m¥/L 20.2°C
Wait until
¡
appears before
the
AR
(signal sound).
The MultiLab P5 controls the quantity of final measurements.
Missing bottles: Overwriting with 000 erases bottle number.
ÃMessen
BSB5 Endmess.
01.03.94/01
Fl.-Nr. E 26
É
ÃMessen
BSB5 Endmess.
01.03.94/01
Fl.-Nr. E00
È
Lost or defective bottle:
Overwrite number 126 with
000.