Fluoride
Fluoride
86
BA61101/08.00/IWa/MLP5-E2
BA61101/08.00/IWa/MLP5-E2
87
Ã
ISE CAL F-
Ñ
C1=009.9 m¥/L
Enter the concentration of
the first standard solution.
: Select figure
: Set figure.
Rinse electrode and temperature probe and immerse
into first standard solution.
ÃC2
ISE CAL F-
U1= 10.8mV
25.4°C
C1= 9.99 m¥/L
Start measurement. Wait
for stable test values.
Ã
ISE CAL F-
C2= 00.00m¥/L
É
Select concentration of the
second calibration standard
solution.
: Select figure
: Set figure.
Rinse electrode and temperature probe and immerse
into second standard solution.
ÃCalDat
ISE CAL F-
U2= -45.3mV
25.4°C
C2=102.11m¥/L
Start measurement. Wait
for stable test values.
U0= -102.3mV
ÂISE F
S= -57.5mV/D
ISE CAL F-
Display of calibration data.
ÂISE F
ISE CAL F-
C1= 9.99m¥/L
C2=102.11m¥/L
Indication of the used
standard solutions.
À
and
Á
switch to-and-fro
between the calibration
data.
Ãmeth.
Âmenu
ÁF-
25.4°C
TP
ÀCal
Back to fluoride menu.