Troubleshooting
Troubleshooting
220
BA61101/08.00/Pla/MLP5-E2
BA61101/08.00/Pla/MLP5-E2
221
ISE N, ISE F and ISE
CalErr TP
ÀCal
Calibration: Slope outside of
admissible range.
Admissible range: -62 ... +62 mV
Possible cause / Action:
- Cable broken
/ Change measuring or reference
electrode or repair meter
- Unsuitable standard solutions
/ Change standard solutions
- Defective or missing
reference electrode
/Change reference electrode
- Wrong or missing preparation of
standard solutions
Measurement: Missing calibration data.
Possible cause / Action:
- RESET performed / Calibrate
- Initial setup / Calibrate
Conductivity
Âmenu >30.00
®µS/cm 25.4°C
nlF 25
Measurement: The measured value is
outside the selected measuring range.
Possible cause / Action:
- Measuring range too small
/ Select bigger measuring range or
automatic selection of measuring range
Âmenu OFL
®µS/cm 25.4°C
nlF 25
Measurement: The measured value is
above the biggest measuring range.
Possible cause / Action:
- The conductivity of the sample solution
is above 500 µS/cm
/ Use meter with a bigger measuring
range
T= 23.3°C
C Ca¨ 0.01mKC¨
(C=1.000/cm)
à C>1.500/cm
Determine cell constant: The cell
constant is above 1.500/cm. With
confirmation the old value remains.
Possible cause / Action:
- Wrong measuring cell / Use a measuring
cell recommended for the meter
- Contaminated or old electrodes
/ Clean electrodes
- Wrong or destroyed control solution
(use potassium chloride solution
c=0.01 mol/l only once)
/ Change control solution