Conductivity measurement
Conductivity measurement
104
BA61101/08.00/IWa/MLP5-E2
BA61101/08.00/IWa/MLP5-E2
105
Setting the temperature function TC
Note: We recommend to use a linear temperature function for the
temperature compensation only with measurements in the
range of ± 10 K around the reference temperature.
Ãmeth.
Âmenu
Á® 25.4°C
nlF 25
Conductivity menu.
TC nlF
®
documentation
¡C /cm
0.609
Menu "additional functions"
for conductivity
measurement.
Tref 25°C
®
C 0.609 /cm
¡TC
nlF
Select Temperature
function nlF.
Âmenu
TC Cal
¡TC
nlF
Select Change of
temperature function.
Âmenu
TC Cal
¡TC %/K
È.ÈÈ
Select Linear temperature
function.
Âmenu
TC Cal
¡TC .00 %/K
È
Set value.
: Select figure
: Set figure.
Setting range
0.00... 3.50 %/K
Âmenu
TC Cal
¡TC %/K
Ê.ËÍ
Confirm input.
Tref 25 °C
®
C 0.609 /cm
¡TC %/K
2.35
Back to menu "additional
functions".
Ãmeth.
Âmenu
Á® 25.4°C
2.35 25
Back to conductivity menu.
Determination of the coefficient for a linear
temperature function: "TC Cal"
To determine a linear temperature function select start and end
temperature so that
•
the set reference temperature Tref is bracketed
•
the end temperature is maximum 10 K and minimum 20 K above
the start temperature.