Conductivity measurement
Oxygen measurement
110
BA61101/08.00/IWa/MLP5-E2
BA61101/08.00/IWa/MLP5-E2
111
Âmeth.
ÁSal
23.5°C
nlF 20
Selection confirmed.
Âmeth.
Sal
23.5°C
nlF 20
36,5
Start measurement.
Salinity measurement
runs continuously.
14564 ÄSO¿
Âmeas
®
¡Sal
Change method.
Further information on conductivity measurement
Fehler! Textmarke nicht definiert. Supervision of conductivity
systems according to ISO 9000;
GLP requirements.
WTW application report no. 395054
Fehler! Textmarke nicht definiert. Conductometric salinity
measurement.
WTW application report no. 993109
Fehler! Textmarke nicht definiert. Temperature compensation of
conductivity values.
WTW application report no. 1194055
Calibration with OxiCal
®-
SL or OxiCal
®
-P
Note: The oxygen probe has to be sufficiently polarized
for calibration!
Polarisation time (Meter switched on, probe
connected):
Ÿ
With initial setup or after cleaning:
approx. 25-50 minutes,
Ÿ
With short interruptions up to 1 h:
approx. 5-20 minutes
Dry oxygen probe approx. ¼ hour befor calibration,
put it into air-calibration beaker OxiCal®and tighten
(protection of temperature adjustment between probe
and air calibration beaker)!
Note: Sponge in the OxiCal® air calibration beaker
must always be moist (not wet!)
Ãmeth.
Âmenu
ÀCal
ÁOº m¥/L 25.4°C
AR
Menu Oxygen
measurement.
HºO-Cal
ÂOº
Sal
¡OxiCal
Select calibration function
OxiCal.
T= 21.6°C
OxiCal
S= 0.983
AR
Start calibration in air
calibration beaker.
AR
: Measured values
"running".
Admissible slope:
S = 0.6... 1.25.