Documentation
Documentation
168
A61101/08.00/IWa/MLP5-E2
BA61101/08.00/IWa/MLP5-E2
169
Photometer
Command
Function dependent on position of selector switch
CMES
Measurement and transmission of sample value
14561 22.02.94 08:45:27 * -0.010 mg/L CN
MLP5 >
CEXT
Measurement and transmission of the extinction value
Ext. 22.02.94 18:33:52 0.012 E 620nm
MLP5 >
CBLA
Measurement and transmission of the sample blank
value . The following concentration measurement
considers the sample blank value.
Erasure of the stored sample blank value by
changement of the method at the selector switch , by
discontinuance of "Remote"-operation or by input of an
own method.
cmes
⇒
14561 22.02.94 08:45:27 * -0.010 mg/L CN
MLP5 >
14561 22.02.94 08:45:30 ! 0.040 mg/L CN
MLP5 >
Command
Function undependent on position of selector switch
CEME
Input of an own method (refer to "Input of own methods
")
MLP5 >
CCLR
Erasure of an own method (refer to "Erasure of own
methods")
MLP5 >
CFIL
Input of data of a 12. filter (refer to "Input of a 12. filter")
MLP5 >
pH undependent on the position of the selector switch!
Command
Function
CPHC
pH-test value transmission. Resolution as set in the
instrument.
pH 22.02.94 13:45:09 7.82 pH 20.7.°C
MLP5 >
INVALID COMMAND
when temp. probe is disconnected
CMVC
mV-test value transmission. Resolution as set in the
instrument.
mV 23.02.94 12:36:21 -48 mV 23.8 °C
MLP5 >
INVALID COMMAND
when temp. probe is disconnected
CISF
Fluoride-test value transmission (if calibration values
available, otherwise INVALID COMMAND)
ISEF 24.02.94 08:02:00 16.3 mg/L F- 23.9 °C
MLP5 >
INVALID COMMAND
when temp. probe is disconnected
CISN
Ammonium-test value transmission (if calibration values
available, otherwise INVALID COMMAND)
ISEN 24.02.94 13:45:00 146 mg/L NH4-N 24.2°C
MLP5 >
INVALID COMMAND
when temp. probe is disconnected
CISE
Ionsensitive test value transmission (if calibration values
available, otherwise INVALID COMMAND)
ISE 24.02.94 13:45:00 146 24.2°C
MLP5 >
INVALID COMMAND
when temp. probe is disconnected