What to do if...
MD 8000 L
86
ba77218e01
11/2018
Error message,
Error
15.3 Conductivity
Error message
OFL, UFL
The measured value is outside the measuring range.
Error message,
Error
15.4 General topics
Sensor symbol
flashes
Cause
Remedy
– Measured value outside the
measuring range
– Use a suitable IDP D.O. sensor
Cause
Remedy
– Sensor contaminated
– Clean the sensor
– Measured temperature value
outside the operating conditions
(display of OFL/UFL instead of a
temperature value)
– Keep to the temperature range for
the test sample
– Defective sensor
– Calibration
– Exchange the sensor cap
– Replace the sensor
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
– Measured value outside the
measuring range
– Use a suitable IDP conductivity sen-
sor
Cause
Remedy
– Sensor contaminated
– Clean the sensor and replace it if
necessary
– Calibration solution not suitable
– Check the calibration solutions
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring system