MD 8000 L
Dissolved oxygen
ba77218e01
11/2018
41
8
Dissolved oxygen
8.1
Measuring
8.1.1 Measuring D.O.
Selecting the
displayed
measured parameter
You can switch between the following displays with
<MODE>
:
D.O. concentration [mg/l]
D.O. saturation [%]
D.O. partial pressure [mbar].
Salinity correction
When measuring the D.O. concentration [mg/l] of solutions with a salt content
of more than 1 g/l , a salinity correction is required. For this, you have to mea-
sure and input the salinity of the measured medium first.
When the salinity correction is switched on, the [Sal] indicator is displayed in
the measuring screen.
1.
Connect the D.O. sensor to the meter.
The D.O. measuring screen is displayed.
2.
If necessary, select the measured parameter with
<MODE>
.
3.
Check or calibrate the meter with the sensor.
4.
Immerse the D.O. sensor in the test sample.
O
2
mg/l
7
.9
2
25
.
0
°C
01.11.2018
08:00
i
USB output
You can switch the salinity correction on or off and enter the salinity
in the menu for calibration and measurement settings (see section
11.3.1 S