Settings
MD 8000 L
66
ba77218e01
11/2018
11.3.2 Enter
cap coefficients
11.3.3
Saturation, local
Irrespective of the height or air pressure, the calibration value is set to 100 %.
The function
Saturation, local
fulfills the EU regulations for the parameter
oxygen saturation [%].
When the
Saturation, local
is enabled the display shows an [L] for the param-
eter oxygen saturation.
D.O. mg/L readings are unaffected by the selection of the
Saturation, local
function.
11.4 Cond measurement settings
11.4.1 Settings for IDP conductivity sensors
Settings
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the
<OK>
key. After completing the settings,
Saturation, local
on
off
Saturation, local
is a procedure that
takes the local air pressure into
account for each saturation mea-
surement.
For details, see section 11.3.3 S
Reset
-
Resets all sensor settings to the
delivery condition (see section
11.7.1 R
The values of the coefficients are provided with the sensor cap.
1.
Change the digit of the highlighted position with
<
><
>
.
2.
Go to the next position with
<
><
>
.
3.
Confirm with
<OK>
when a coefficient is completely entered.
Menu item
Possible
setting
Explanation