9.3.4 Calibration data
The calibration data can be displayed and then output to the interface.
Displaying the cali-
bration record
The calibration protocol of the last calibration is available in the menu
Calibra-
tion
/
Calibration record
. To open it in the measured value display, press the
<CAL__>
key.
The calibration records of the last 10 calibrations are available in the menu
Cali-
bration
/
Calibration data storage
/
Display
. To open the
Calibration
menu in the
measured value display, press the
<OK>
key.
5.
Start the measurement with
<OK>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
6.
Wait for the end of the measurement with stability control ([HOLD][AR])
status indicator or
take over the calibrated value with
<OK>
.
The calibration record is displayed and output to the interface.
7.
Switch to the measured value display with
<OK>
.
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item
Setting/
function
Explanation
Calibration
/
Calibration data
storage
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibration
record to the interface with
<F2>
/
[USB output]
.
Output all calibration records to
the interface with
<F2__>
[USB
output]
.
Quit the display with
<F1>
/
[Zurück] or
<OK>
.
Switch directly to the measured
value display with
<MODE>
.
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the calibration data storage
to the interface