![VWR MD 8000 L Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/vwr/md-8000-l/md-8000-l_operating-manual_1057579037.webp)
MD 8000 L
pH value
ba77218e01
11/2018
37
Calibration record
(
USB output
)
Error
Error
<-30
or
>+30
<-62.0
or
> -50,0
Error elimination (see section 15 W
Display
Calibration
record
Zero point
[mV]
Slope [mV/pH]
MD 8000 L
Ser. no. 11292113
CALIBRATIONpH
01.11.2018 15:55
Ser. no. 10501234
TEC
Buffer 1
10.00
Buffer 2
7.00
Buffer 3
4.00
Voltage 1
-177.0 mV
Voltage 2
3.0 mV
Voltage 3
184.0 mV
Temperatur 1
24.0 °C
Temperatur 2
24.0 °C
Temperatur 3
24.0 °C
Slope
-60.2 mV/pH
Asymmetry
4.0 mV
Sensor
+++
etc...