![VWR MD 8000 L Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/vwr/md-8000-l/md-8000-l_operating-manual_1057579056.webp)
Turbidity measurement (VisoTurb
®
900-P)
MD 8000 L
56
ba77218e01
11/2018
Selecting the
displayed
measured parameter
You can switch between the following displays with
<M>
:
Turbidity [FNU]
Turbidity [NTU]
Freezes the mea-
sured value (HOLD
function)
With the HOLD function, you can freeze the current measured value. The dis-
played measured value stops changing until you switch the HOLD function off.
Stability control
(AutoRead )
The stability control function (
AutoRead
) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values. The display of the measured parameter flashes until
a stable measured value is available.
You can start the
Stability control
manually at any time, irrespective of the set-
ting for automatic
Stability control
System
menu.
TRB
F
NU
7
.9
2
01.02.2014
08:00
i
USB output
1.
Freeze the measured value with
<HOLD>
.
The [HOLD] status indicator is displayed.
If the HOLD function is active, you can, e.g. start a manual mea-
surement with stability control.
2.
Release the frozen measured value again with
<HOLD>
.
The HOLD function is switched off.
The [HOLD] status display disappears.
1.
Freeze the measured value with
<HOLD>
.
The [HOLD] status indicator is displayed.