11.1.2 Buffer sets for calibration
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during the calibration.
Brand names or trade names are trademarks of their respective owners protected
by law.
No.
Buffer set *
pH values
at
1
TEC
Technical buffers
2.00
4.00
7.00
10.00
20°C
2
AnyCal
Any
Any
3
NIST/DIN
DIN buffers according to DIN
19266 and NIST Traceable
Buffers
1.679
4.006
6.865
9.180
12.454
25 °C
4
TEC 2
Technical buffers
2.000
4.010
7.000
10.011
25 °C
5
Merck 1*
4.000
7.000
9.000
20°C
6
Merck 2 *
1.000
6.000
8.000
13.000
20°C
7
Merck 3 *
4.660
6.880
9.220
20°C
8
Merck 4 *
2.000
4.000
7.000
10.000
20°C
9
Merck 5 *
4.010
7.000
10.000
25 °C
The buffers are selected in the menu, pH /
<OK>
/
Calibration
/
Buffer
(see section 11.1.1 S