![VWR MD 8000 L Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/vwr/md-8000-l/md-8000-l_operating-manual_1057579065.webp)
MD 8000 L
Settings
ba77218e01
11/2018
65
tings, switch to the measured value display with
<MODE>
.
Default settings are printed in
bold
.
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record of
the last calibration.
Calibration
/
Calibration data
storage
/
Display
-
Displays the last calibration
records.
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the calibration data stor-
age to the interface
Calibration
/
Calibration interval
1 ...
180
...
999 d
Calibration interval
for the D.O.
sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring screen.
Calibration
/
Comparison meas.
on
off
Enables to adjust the measured
value with the aid of a comparison
measurement, e.g. Winkler titra-
tion.
For details, see section 8.2 C
Calibration
/
cap
coefficients
K1
...
K5
KC
Enter the cap coefficients here after
exchanging the sensor cap.
For details, see section 11.3.2
E
Salinity
/
Sal correc-
tion
on
off
Manual salt content correction for
concentration measurements.
Salinity
/
Salinity
0.0
... 70.0
Salinity or salinity equivalent for the
salt content correction.
This menu item is only available if
the manual salinity correction is
switched on.
Resolution
0.1
1
Set a high or low resolution.
The setting of the resolution is sep-
arately stored for each measured
parameter.
Содержание MD 8000 L
Страница 93: ...MD 8000 L Glossary ba77218e01 11 2018 93...
Страница 97: ......