MD 8000 L
Dissolved oxygen
ba77218e01
11/2018
43
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
8.1.2 Measuring the temperature
For reproducible D.O. measurements, it is essential to measure the tempera-
ture of the test sample.
D.O. sensors measure the temperature with a temperature sensor integrated
in the sensor.
8.2
Calibration
8.2.1 Why calibrate?
D.O. sensors age. Aging changes the zero point and slope of the D.O. sensor.
As a result, an inexact measured value is displayed. Calibration determines
and stores the current values of the zero point and slope.
8.2.2 When to calibrate?
When the calibration interval has expired
When your accuracy requirements are especially high
Routinely within the framework of the company quality assurance
You can prematurely terminate the
Stability control
function manu-
ally with
<OK>
at any time. If the
Stability control
function is pre-
maturely terminated, the current measurement data are output to
the interface without the AutoRead info.
You can switch off the beep (see section 11.6 S
3.
Using
<OK>
, start a further measurement with stability control.
or
Release the frozen measured value again with
<HOLD>
or
<MODE>
.
The [AR] status display disappears. The display switches back to the
previous indication.
Measured parameter
Time interval
Stability in the time
interval
D.O. concentration
20 seconds
∆
: better than 0.03 mg/l
D.O. saturation
20 seconds
∆
: better than 0.4 %
Temperature
15 seconds
∆
: better than 0.5 °C