Dissolved oxygen
MD 8000 L
42
ba77218e01
11/2018
Air pressure
correction
(
Saturation, local
function)
The integrated air pressure sensor of the MD 8000 L measures the current air
pressure. During calibration, the air pressure correction function is automati-
cally activated. While the parameter oxygen saturation [%] is displayed, the air
pressure correction is applied if the
Saturation, local
function is active.
Stability control
(AutoRead)
& HOLD function
The stability control function (
AutoRead
) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic
Stability control
is switched off.
Irrespective of the setting for automatic
Stability control
(see section 11.6.3
, page 70) in the
System
menu, you can start a
measurement with
Stability control
manually at any time.
To start the
Stability control
function manually, the HOLD function must be
active.
Hold function
Stability control
The air pressure correction (function
Saturation, local
) is switched
on or off in the menu for calibration and measurement settings (see
section 11.3.1 S
1.
Freeze the measured value with
<HOLD>
.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the HOLD function and the
Stability control
func-
tion with
<HOLD>
or
<MODE>
at any time.
2.
Using
<OK>
, activate the
Stability control
function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the mea-
sured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears, the display of the mea-
sured parameter stops flashing and a beep sounds.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.