6.2.5 Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
6.2.6 Calibration data
Displays the calibra-
tion data
The calibration protocol of the last calibration is available in the menu
Calibra-
tion
/
Calibration record
. To open it in the measured value display, press the
<CAL__>
key.
The calibration records of the last 10 calibrations are available in the menu
Cali-
bration
/
Calibration data storage
/
Display
. To open the
Calibration
menu in the
measured value display, press the
<OK>
key.
Calibration
Determined
values
Displayed calibration data
1-point
Asy
Zero point =
Asy
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point
Asy
Slp.
Zero point =
Asy
Slope =
Slp.
3-point to 5-
point
Asy
Slp.
Zero point =
Asy
Slope =
Slp.
The calibration line is calculated by
linear regression.
You can display the slope in the units, mV/pH or % (see section
11.1.1 S
The calibration record is automatically transmitted to the interface
after calibrating.