Settings
MD 8000 L
62
ba77218e01
11/2018
11 Settings
11.1 pH measurement settings
11.1.1 Settings for pH measurements
Settings
The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required mea-
sured parameter in the measured value display and press the
<OK>
key. After
completing the settings, switch to the measured value display with
<MODE>
.
Default settings are printed in
bold
.
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record of the last calibration.
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the last calibration records (max. 10)
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the calibration data storage to the interface
Calibration
/
Buffer
TEC
AnyCal
NIST/DIN
...
Buffer sets to be used for pH calibration.
More buffers and details: see section 11.1.2 B
Calibration
/
Single-
point calibration
yes
no
Quick calibration with 1 buffer
Calibration
/
Calibration interval
1 ...
7
... 999
d
Calibration interval
for the IDPpH sensor (in days).
The meter reminds you to calibrate regularly by the flashing
sensor symbol in the measuring screen.
Calibration
/
Unit for
slope
mV/pH
%
Unit of the slope.
The % display refers to the Nernst slope of
-59.2 mV/pH (100 x determined slope/Nernst slope).
pH resolution
0.001
0.01
0.1
Resolution of the pH display
mV resolution
0.1
1
Resolution of the mV display
Reset
-
Resets all sensor settings to the delivery condition (see sec-
tion 11.7.1 R