The meter can also use manual salinity correction for dissolved oxygen readings
reported as mg/L. The manual salinity correction factor must be entered as ppt
(parts per thousand).
Conductivity
at 20 °C
(mS/cm)
Salinity
Correction
Value (ppt)
Conductivity
at 20 °C
(mS/cm)
Salinity
Correction
Value (ppt)
Conductivity
at 20 °C
(mS/cm)
Salinity
Correction
Value (ppt)
5
3
20
13
35
25
6
4
21
14
36
25
7
4
22
15
37
26
8
5
23
15
38
27
9
6
24
16
39
28
10
6
25
17
40
29
11
7
26
18
42
30
12
8
27
18
44
32
13
8
28
19
46
33
14
9
29
20
48
35
15
10
30
21
50
37
16
10
31
22
52
38
17
11
32
22
54
40
18
12
33
23
56
42
19
13
34
24
This table was calculated from the International Oceanographic Tables, Vol. 1, National Institute of Oceanography
of Great Britain, Womley, Godaming, Surrey, England and Unesco, Paris 1971.
Conductivity Setup Menu Features
Temperature Compensation and Reference Temperature
The sympHony conductivity meters have the ability to use a temperature
compensation feature that calculates and displays the conductivity measurements
at a reference temperature of 15 °C, 20 °C or 25 °C. The temperature compensation
can be set as linear for most aqueous samples, non-linear for ultra pure and low
ionic strength samples or off for non-temperature compensated conductivity
measurements. The closer the sample temperature is to the selected reference
temperature, the more accurate the conductivity measurement will be, especially if
the temperature compensation coefficient is estimated or inaccurate.