12
sympHony
™
Meter User Guide
Setup Menu Table
The following table is for the complete line of sympHony meters. Some sympHony
meters may not include all of the options listed in this table. Refer to
Appendix A
for additional information on the special setup menu features.
Top Line
Middle Line
Bottom Line
Setup Menu Description (default setting, method specific)
PH
rES
0.1, 0.01, 0.001
pH measurement resolution
(0.01, yes)
PH
bUF
USA, EUr0
pH buffer set for automatic buffer recognition during calibration,
USA buffers are 1.68, 4.01, 7.00, 10.01, 12.46 and EUrO buffers are
1.68, 4.01, 6.86, 9.18
(USA, yes)
ISE
rES
1, 2, 3
ISE measurement resolution in significant figures
(1, yes)
ISE
UnIt
m, mgL, PEr, PPb, n0nE
ISE measurement units
(PPb, yes)
ISE
rAng
L0w, HIgH
ISE concentration range for calibration stability criteria
(HIgH, yes)
ISE
nLIn
AUt0, 0FF
ISE automatic blank correction for low-level calibration
(AUt0, yes)
C0nd
tC
0FF, LIn, nLF
Conductivity temperature compensation type, LIn is for linear, nLF is
for non-linear pure water samples
(LIn, yes)
C0nd
C0EF
0.0 to 10.0
Conductivity temperature compensation coefficient in % change in
conductivity per °C, appears if LIn was selected for tC
(2.1, yes)
C0nd
tdSF
0.00 to 10.0
Conductivity TDS factor value
(0.49, yes)
C0nd
CELL
0.001 to 199.0
Conductivity default cell constant value for automatic conductivity
calibration mode
(0.475, yes)
C0nd
trEF
15, 20, 25
Conductivity reference temperature
(25, yes)
C0nd
tyPE
Std, 1, 2, 3, 4, 5, 6, 7
Conductivity cell type and selectable range
(Std, yes)
d0
rES
0.1, 1 % sat
DO % saturation measurement resolution
(0.1, yes)
d0
rES
0.01, 0.1 mg/L
DO mg/L measurement resolution
(0.01, yes)
d0
bAr
AUt0, mAn
DO barometric pressure compensation type
(AUt0, yes)
d0
PrES
450.0 to 850.0
DO manual barometric pressure compensation value, appears if
mAn was selected for bAr
(760.0, yes)
d0
SAL
AUt0, mAn
DO salinity correction type
(AUt0, yes)
d0
SALF
0 to 45
DO manual salinity correction value, appears if mAn was selected
for SAL or a DO meter without a conductivity mode is used
(0, yes)
d0
CALt
AIr, H20, mAn, SEt0
DO calibration type
(AIr, yes)
dUE
PH
0 to 9999
pH calibration alarm value in hours, 0 is off
(0, yes)
dUE
0rP
0 to 9999
ORP calibration alarm value in hours, 0 is off
(0, yes)
dUE
ISE
0 to 9999
ISE calibration alarm value in hours, 0 is off
(0, yes)
dUE
C0nd
0 to 9999
Conductivity calibration alarm value in hours, 0 is off
(0, yes)
dUE
d0
0 to 9999
DO calibration alarm value in hours, 0 is off
(0, yes)
Meter Setup