35
sympHony
™
Meter User Guide
Chapter X
Data Archiving and Retrieval
Datalog and Calibration Log
The sympHony meters have a 200 point datalog that includes the last ten
calibrations that were successfully performed on the meter.
Datalog Deletion Setting
The datalog deletion setting determines if the meter will automatically delete
the datalog after it is downloaded to a printer or computer and if the meter will
overwrite the datalog points when the datalog is full. If the datalog deletion setting
is set to
, the meter will automatically delete the datalog after the datalog
is downloaded to a printer or computer. The meter will also display an error 038
message when all 200 datalog points are filled and the datalog must be downloaded
to a printer or computer to clear the error message. If the datalog deletion setting
is set to
, the meter will overwrite the oldest datalog point when all 200 datalog
points are filled.
1. In the measurement mode, press
.
2. Press
/
until
is displayed on top line.
3. Press
to accept the selection and move the arrow icon to the middle line
and press
/
until
is displayed.
4. Press
to accept the selection and move the arrow icon to the bottom line
and press
/
until
or
is displayed.
5. Press
to accept the selection and move the arrow icon to the top line.
6. Press
to save the setup option and return to measurement mode.
Note:
If the datalog is not required, set the datalog deletion setting to
to
prevent the error 038 (datalog full) message.