b. Direct calibration – When the
µS/cm
or
mS/cm
icon stops flashing, the
meter will display the actual conductivity value read by the probe. To
change the conductivity value, press
until the first digit to be changed
is flashing, press
/
to change the value of the flashing digit and
continue to change the digits until the correct conductivity value of the
standard at the measured temperature is displayed. Once the value is set,
press
until the decimal point is in the correct location.
5. Press
to proceed to the next calibration point, rinse the conductivity probe
with distilled or deionized water, place it into the next conductivity standard and
repeat step 4a / 4b or press
to save and end the calibration.
6. The cell constant will be displayed in the main field and the meter will
automatically advance to the measurement mode.
Conductivity Technique