EVERY TWO YEARS
•
After installation and following significant repairs, and in any case every two years, use an electrical safety testing
instrument to ensure that the device has maintained its safety characteristics (ref. EN ISO 60601-1). Use as a
reference the data indicated in the inspection report enclosed with the machine instruction manual, in the
chapter "
Technical Machine Specifications
".
When checking, all devices connected to the equipment have to be under tension and turned on.
•
Note:
For keeping the expected safety standards, a general check of the equipment should be carried out at
least every two years. For this, use the installation reference card indicate in the chap. “
Installation Inspection
”.
Additional maintenance operatio
Additional maintenance operations are required in the presence of any devices not manufactured
by VITALI. For intervention procedures, see the corresponding user manuals (i.e., dynamic
amalgam separator, water treatment devices, etc.).
Replace the filter cartridge "
A
".
Replacing the air unit filter
A