13
T 5
Assembling the instrument table arm
Proceed as follows to assemble the instrument table arm:
1) Insert all of the cables and hoses “
E
” from the instrument table arm through the hole “
D
”
2) Secure the tray table to the dental unit bracket with the four M8 screws “
A
” provided, level the arm with the three
M10 dowel pins “
B
” located on side "
C
".
3) Level using surface "
F
" as a reference allowing for possible bending and tighten the four Allen M8 screws "
A
".
4) Tighten the three M10 dowels "
B
" to avoid losing the leveling adjustment.
5) After installation, apply protection guard "
M
" on the bracket "
C
" by means of the two M5 screws.