63
T 5
Adjusting the instrument table support panel friction
This operation should be carried out whenever the weight of instrument table changes (for instance, by adding
or removing an instrument), or whenever it may become necessary.
Remove the guard “
A
” in the direction shown by the arrows.
To achieve the proper friction, tighten or loosen the nuts “
B
”. The lifting force of the instrument table is not
adjustable, since it is determined by the force exerted by the gas springs “
C
”.
After adjusting, re-mount the housing “
A
” in the opposite direction to one shown by the arrows.
☞
NOTE:
The parallel between axes “
E
” and “
F
” the stops “
G
” and “
H
” are adjusted at our
factory, and require
NO further adjustment.
☞
NOTE: NEVER unscrew the screws “L”!