11
T 5
Mounting the dental unit bracket on the chair
Proceed as follows:
1) Insert the tube bundle “
A
” (bundle of electrical cables clamped to the ribbed grey suction hose id. 25 and the
sheath containing the air and water intake hoses) through the hole “
D
” in the bracket “
C
”.
2) Insert the hose "
B
" (transparent Armovin water drainage hose diam. 20x27).
3) Attach the bracket in area "
E
" through the holes "
F
" using the three M10 Allen screws.
4) Level using surface "
G
" as a reference by means of the four M10 dowels "
L
" and tighten the three M10 Allen
screws.
5) After fastening the bracket, cut the plastic clamps on the bundle of hoses "
A
".
6) Fit the closing guard "
H
" on the seat guard "
I
" and fasten by means of the countersunk screws M6 supplied.
7) Cover the M6 screws using the two white covers supplied.