T 5
40
STANDARD
CONFIGURATION
1
ON
2
3
4
5
6
7
8
Dip-Switch no.
ON
OFF
1
Enables the polymerizing lamp slot.
Disables the polymerizing lamp slot.
Affects output "
A
" (PIN 1 e 2) on M7
(24 V AC-150 VA max).
2
Enables the saliva ejector slot.
Disables the saliva ejector slot.
Affects output "
B
" (PIN 3 and 4) on M7
(24 V AC-150 VA max).
3
Enables the anti-glare function.
Disables the anti-glare function.
The operative lamp shuts off
while the chair is moving.
4
The tumbler filling key has no other
The tumbler filling key also starts
functions.
the bowl flush.
5
Complete cycle by bowl flush key
Partial cycle by bowl flush key
6
NOT USED
NOT USED (standard position)
7
NOT USED
NOT USED (standard position)
8
The tumbler filling and bowl flush
The tumbler filling and bowl
solenoid valves work.
flush solenoid valves do not work.
1
2 3
4 5
6 7 8
ON
F10
F11
F12
CN23
CN43
CN35
F13
F14
F15
F16
CN37
CN22
TRIM4
TRIM2
TRIM1
CN29
SW1
CN32
CN36
CN39
CN40
CN38
CN30
TRIM 3
M7
M8
CN27 CN27
CN41
CN42
CN31
CN44
CN26
CN25
CN24
CN34
MICROPROCESSOR
Caution!
Incorrect positioning of the dip-switches
3-4-5-6-7-8 may cause the device to
malfunction
Dental unit card
Replacing the dental unit card fuses
Below is a table showing the errors that may occur when the fuses listed are tripped. See the illustration on the
previous page for fuse locations.
PROTECTED
VALUE
ERRORS THAT OCCUR
PARTS
AND TYPE
WHEN TRIPPED
F10
Dynamic amalgam
F. 6,3A (5x20)
The dynamic amalgam separator does not work.
separator power supply
RAPID
F11
Assistant/dentist
F. 6,3A (5x20)
The instruments do not work.
console power supply,
RAPID
The x-ray viewer does not work. The tumbler
water unit solenoid valves
filling and bowl flush solenoid valves do not work.
F12
Tumbler water heater, suction
F. 6,3A (5x20)
The assistant call button does not work.
control, assistant call button,
RAPID
The tumbler water heater does not work.
x-ray viewer power supply
The x-ray viewer does not work.
F13
Polymerizing lamp power
F. 5A (5x20)
The polymerizing lamp does not work.
supply
RAPID
F14
Saliva ejector solenoid valve
F. 5A (5x20)
The saliva ejector does not work.
power supply
RAPID
F15
Dental unit card
F. 1A (5x20)
Only the instruments work.
power supply
RAPID
F16
Operative lamp
F. 6,3A (5x20) RAPID
The operative lamp does not work.
power supply
(with halogen lamp EDI)
F. 1A (5x20) RAPID
(with LED lamp ALYA)
LABEL
Setting the dental unit card dip-switches