41
T 5
To remove the upper table guard, proceed as follows:
1) Remove the instrument support "
A
"
2) Disassemble the fixing bushes of the instrument
cords "
B
"
3) Move the instrument cords outward, as shown in
the fig. 1
4) Unscrew the 7 screws "
C
", as shown in the fig. 2
5) Disconnect the connectors "
M
" and "
L
" from control
keyboard (fig. 3).
Follow the above steps in reverse order to reassemble.
15 ACCESSING THE INTERNAL PARTS OF THE INSTRUMENT TABLE
Fig. 1
Fig. 2
NOTE:
To ensure the perfect matching,
check the correct positioning of both table
guards before fastening the 7 screws "C".
L
M
Fig. 3
C
C
C C
C
C
C
Rm
E
1
2
3
4
5
6
7
8
9
10
Rm
E
1 2
3 4
5 6
7 8
9 1
0
B
A