23
T 5
Setting the FLARES chair movement Dip-Switches
Should you wish to move the chair during installation before you have installed the water unit, you must adjust the
switch SW1 on the card "
D
" as shown. To access this, raise the derivative housing "
A
", unscrew the two holding
screws "
B
" and remove the protective housing "
C
".
WARNING!!
Cut off electrical power before removing the housing!
Removal of the connector block housing provides access to parts that are
live even when the power switch is turned off.
A
Should you wish to move the chair during installation before you have installed
the water unit, you must adjust the switch SW1 as shown.
Warning!
Note the position of the dip-switches before making any changes.
1
ON
2
3
4
5
6
7
8
SW1
CN14
CN17
SP3
CN13
CN12
CN11
1
ON
2
3
4
5
6
7
8
Caution!
Incorrect positioning of the
dip-switches
3-4-5-6-7-8 may cause the device
to malfunction
Warning!
After installing the water unit, set the dip-switch to the position
shown in the figure.
Note:
The position of the dip-switch varies depending on the equipment
configuration (see table below).
1
ON
2
3
4
5
6
7
8
STANDARD CONFIGURATION
FLARES
Dip-Switch no.
OFF
ON
1
Not used
Not used
2
The "mechanical memory" key has the
The "mechanical memory" key has
function of a "rinse" key
its standard function
3
Starting an instrument prevents the
Chair operation is independent
chair from moving
from instrument use
4
The instruments work even while the
Moving the chair prevents the
chair is moving
instruments from being used
5
Enables the "pantographic arm safety"
Tripping the "pantographic arm safety"
microswitch (
CN11
)
microswitch does not affect chair
operation
6
Enables the "backrest safety"
Tripping the "backrest safety"
microswitch (
CN12
)
microswitch does not affect chair
operation
7
Enables the "seat safety" microswitch
Tripping the "seat safety" microswitch
(
CN13
)
does not affect chair operation
8
Enables the "assistant arm safety"
Tripping the "assistant arm safety"
microswitch (
CN14
)
microswitch does not affect chair
operation.
Chair version definition
SOFTWARE
JUMPER
RELEASE
CN17
SP3
Basic
Open
Open
With "mechanical"
memory
Open
Closed
With memories
Closed
Open
NOTE:
In case of Versions with Memories, additional devices
are necessary for the correct operating !
D
C
B
B
Содержание FLARES
Страница 1: ...T5 Re Rm Technician EN...
Страница 2: ......
Страница 10: ......
Страница 81: ...77 T5 27 WIRING DIAGRAMS...
Страница 82: ......
Страница 84: ......
Страница 86: ......