65
T 5
Assembling the instrument table arm stop
A
B
C
D
1) Open the upper table housing as described in the
chapter "
Accessing the internal parts of the
instrument table
".
2) Remove the arm by unscrewing the four self-
threading screws "
A
".
3) Insert the stopping cam "
B
" into its slot.
4) Insert the spring "
C
" as illustrated.
5) Fasten the cams with the self-threading screw "
D
"
provided, taking care to leave enough room to
allow the cam to move properly.
Assembling the instrument table arm
1) Open the upper table housing as described in the
chapter "
Accessing the internal parts of the
instrument table
".
2) Remove the 4 screws “
A
”.
3) Insert the arm as illustrated.
4) Finally, use the 4 self-tapping screws “
B
” provided.
A
A
A
A
B