55
T 5
Electronic foot control setting and configuration
Move the dip-switch 2 to On (the LED on the card flashes quickly at equal intervals).
2
•
Move the lever all the way to the right, then slowly release it until it returns to resting position.
3
•
Move the lever all the way to the left, then slowly release it until it returns to resting position.
4
•
Move the dip-switch 2 to Off. (the LED on the card flashes slowly in pulses).
5
•
Place a volt meter on pins 7 and 2 of CN82, and make sure that the voltage reading with the lever all the way
to the right is the same as the reading with the lever all the way to the left. If so, the setting is complete and
you may proceed with closing the foot control. If not, repeat the procedure (steps 1 through 4), moving the
lever only to the side on which the volt meter measured a lower value.
NOTE
: When the LED flashes, alternating long lit phases with short off breaks, you must check the serial connection
(CN80) or the setting.
No. dip-switches S1
Off
On
1
Not used
Not used
2
Acquire setting
Min/max foot control lever setting
3
Spray by moving the lever
Spray by moving the lever
to the right
to the left
4
Not used
Not used
5
Not used
Not used
6
Not used
Not used
4
5
6
The position of the dip-switches
4-5-6
S1
determines the function of the push-button placed on the foot control
(6). The following table indicates the functions that can be activated.
OFF
OFF
ON
“Rinsing” position.
OFF
ON
ON
Automatic chair reset.
ON
OFF
OFF
Reversal of the sense of rotation of the micromotor.
ON
ON
ON
Timed tumbler filling.