47
T 5
Remove the derivative housing as described in the chapter "
Accessing the internal parts
", par. "
Removing the
derivative housing
".
1
•
Main air inlet (blue diam. 8 hose).
2
•
Air filter with condensation drainage valve.
3
•
Air pressure regulator for the headrest bayonet brake and air pressure of the pressurized bottle
(optional).
4
•
Air outlet for the headrest bayonet brake and air outlet for the pressurized bottle (white hose diam 4 - 3 bar).
5
•
Instrument table air pressure regulator.
6
•
Instrument table air output (blue diam. 6 hose). (to adjust based on the type of turbine used).
7
•
Main water inlet (green diam. 8 hose).
8
•
Water filter.
9
•
Instrument table water pressure regulator.
10
•
Instrument table water outlet (green diam. 4 hose - 2 bar).
11
•
Anti-buildup device (optional).
12
•
Tumbler, bowl and saliva ejector water outlet solenoid valve (green diam. 8 hose).
13
•
Chip air, instrument spray and syringe air pressure regulator.
14
•
Chip air, instrument spray and syringe air output.
15
•
Pressurized bottle safety valve.
16
•
Water prefilter.
☞
NOTE:
Removing the derivative housing allows access to parts that remain live even when
the main power switch is shut off.
18 HYDROPNEUMATIC ADJUSTMENTS
7
16
11
7
12
8
4
6
10
1
15
2
14
3
13
5