G4 Multi Channel Force Instrument
6-5
Main menu
To reach the main menu, press the ‘Info’ button on the instrument front panel.
Navigating in a menu is done with arrow keys or by tapping on the display on specific
choices e.g. ‘Parameter set-up’.
To open a sub menu, e.g.
Parameter set-up, select the
sub menu with arrow keys
followed by pressing the
‘Enter key’, or by tapping on
the name.
The instrument Main menu can be opened without interrupting the measuring operation.
In the Main menu it is possible to select between the following sub menus:
Levels:
Viewing and editing of level value for each configured Level Supervision
function. Note that configuration of Level Supervision functionality is done in the
‘Parameter Set-up’ menu.
Clock Set-up:
Used to set the instrument clock and date. Time and Date formats are
set-up with parameters trough ‘Parameter set-up’ menu.
Parameter Set-up:
Access to the G4 instrument parameter set-up menu system. See
chapter ‘Set-up’ for more details on setup.
System Information:
Displaying system information for hardware and software in the
instrument. Program versions, database version, serial numbers and software versions
for all electronic modules can be read here. Info about the Ethernet connection is also
available.
Maintenance:
Includes a number of functions for maintenance purposes. The available
functions are Diagnostics, File Handling, Create Backup, Restore Backup, Set Default
Values and Program Upgrade.
See chapter ‘Maintenance’ for a detailed description of the functions.
Network Configuration:
The Network Configuration menu consists of the IP
Configuration and the Server Configuration menus. The IP Configuration menu is used
to set, manual/auto IP address assignment, IP address etc. The Server Configuration
menu is used to enable/disable the FTP Server and the WEB Server of the instrument.
In this menu is also the password for the servers set.
Note that Ethernet configuration is not done with set-up parameters and is for that
reason not saved with the instrument back-up function.
To return to the operating menu press the ‘Esc’ key on the front panel or the ‘Esc’
button on screen.
Main menu.
Содержание G4-DT
Страница 1: ...G4 Multi Channel Force Instrument Program version 101 1 0 0 Technical Manual PM DT HE types GB ...
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 36: ...Technical Manual 2 18 ...
Страница 45: ...G4 Multi Channel Force Instrument 4 5 ...
Страница 76: ...Technical Manual 4 36 ...
Страница 82: ...Technical Manual 5 6 ...
Страница 92: ...Technical Manual 6 10 ...
Страница 138: ...Technical Manual 7 46 ...
Страница 158: ...Technical Manual 8 20 ...
Страница 164: ...Technical Manual 9 6 ...
Страница 170: ......
Страница 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Страница 172: ...Technical Manual ...
Страница 173: ......